Evaluation of a Point-of-Care Device for Hemoglobin and Hematocrit Measurement in Infants Compared with a Standard Laboratory Analyzer
DOI:
https://doi.org/10.46966/msjar.v7i1.354Keywords:
Hemochroma PLUS, Point-Of-Care Testing, Hemoglobin, Hematocrit, Anemia, InfantsAbstract
Background: Rapid detection of anemia is critical for prompt management in infants. Point-of-care testing (POCT) devices provide immediate results and may be particularly useful in low-resource settings. This study evaluated the Hemochroma PLUS hemoglobinometer for hemoglobin (Hb) and hematocrit (HCT) measurement in infants, compared with the Genex Count 60 laboratory analyzer. Methods: A prospective diagnostic accuracy study was conducted at Al-Zahraa Teaching Hospital for Obstetrics and Pediatrics. Capillary blood samples were collected from 33 infants (age 3–12 months) and analyzed using both Hemochroma PLUS and the reference analyzer. Accuracy, precision, sensitivity, specificity, positive predictive value (PPV), and negative predictive value (NPV) were calculated. Bland–Altman analysis and coefficient of variation (CV) were used to assess agreement and precision. Results: Hemochroma PLUS showed good agreement with the reference analyzer for Hb (mean bias 0.75 g/dL, 95% limits of agreement –1.37 to 3.25 g/dL) but overestimated HCT (mean bias 6.1%, 95% LOA –5.69 to 17.9%). Coefficient of variation was similar between devices (Hb CV ~0.16–0.18). Sensitivity and specificity for Hb-based anemia detection in infants >6 months were 75% and 100%, respectively, with PPV 100% and NPV 93%. In infants 3–6 months, sensitivity decreased to 25% while specificity remained 100%. Conclusion: Hemochroma PLUS is a rapid, economical, and convenient POCT device suitable for hemoglobin measurement in infants older than six months. While reliable for detecting anemia in older infants, confirmatory laboratory testing is recommended for younger infants, and caution is advised for HCT-based assessments.
Downloads
Published
How to Cite
Issue
Section
License
Copyright (c) 2026 Ali M. Kadhim

This work is licensed under a Creative Commons Attribution 4.0 International License.
In submitting the manuscript to the Medical Science Journal for Advance Research, the authors certify that:
- They are authorized by their co-authors to enter into these arrangements.
- The work described has not been formally published before, except in the form of an abstract or as part of a published lecture, review, thesis, or overlay journal.
- That it is not under consideration for publication elsewhere,
- The publication has been approved by the author(s) and by responsible authorities – tacitly or explicitly – of the institutes where the work has been carried out.
- They secure the right to reproduce any material that has already been published or copyrighted elsewhere.
- They agree to the following license and copyright agreement.
License and Copyright Agreement
Authors who publish with Medical Science Journal for Advance Research agree to the following terms:
- Authors retain copyright and grant the Medical Science Journal for Advance Research right of first publication with the work simultaneously licensed under Creative Commons Attribution License (CC BY 4.0) that allows others to share the work with an acknowledgment of the work's authorship and initial publication in this journal.
- Authors can enter into separate, additional contractual arrangements for the non-exclusive distribution of the Medical Science Journal for Advance Research published version of the work (e.g., post it to an institutional repository or edit it in a book), with an acknowledgment of its initial publication in this journal.
- Authors are permitted and encouraged to post their work online (e.g., in institutional repositories or on their website) before and during the submission process, as it can lead to productive exchanges, as well as earlier and greater citation of published work.